Low-frequency pulsed currents are trains of short electrical pulses, up to 1,000 per second, that fire nerves and muscles. This page covers how they are classified, the parameters that describe them, what they do, and how the main currents of the professor's course are applied and dosed, from pulsed galvanic, tetanic and exponential currents to the diadynamic (Bernard) currents, high-voltage pulsed current and microcurrent.

At a glance

  1. Low-frequency currents are pulses at up to 1,000 Hz. Each pulse long and strong enough fires a nerve once, so these currents speak the nerves' own language.
  2. They are classified by polarity (monophasic or biphasic), pulse shape (rectangular, triangular, trapezoidal, sinusoidal, semi-sinusoidal, exponential), modulation (none, bursts, amplitude, frequency) and clinical use.
  3. Intensity decides sensory or motor; frequency decides the pattern: around 80 to 150 Hz for pain relief at sensory level, 1 to 50 Hz for stimulation, 50 Hz for a smooth (tetanic) contraction.
  4. The professor's currents: pulsed (impulse) galvanic current, tetanic (faradic) current at 50 Hz, exponential current for denervated muscle, and the electroanalgesic currents (TENS, high-voltage pulsed current, microcurrent, diadynamic).
  5. Diadynamic (Bernard) currents are rectified 50 Hz mains half-waves in five forms: DF (100 Hz, strongest but shortest pain relief), MF (50 Hz), CP and LP (alternations of the two, longer relief) and RS (50 Hz with pauses, for muscle).
  6. Pain relief comes from the spinal gate (large touch fibres damp pain fibres), descending inhibition and endorphins; muscle effects from contraction.
  7. Monophasic currents carry a direct-current component and can burn like galvanization. Contraindications are those of every current through the body; evidence for most traditional currents is limited.

Classification

Electrotherapy with pulses works because nerves fire in pulses too. A pulsed current sends a brief push, a pause, another push. If each push is strong enough and long enough, it fires the nerve once. How often, how strong and in what shape the pushes come decides what the patient feels and what the muscle does.

By polarity

TypeWhat it isConsequence
Monophasic (unidirectional)Every pulse flows the same wayA direct-current component: polarity effects, electrolysis, risk of a chemical burn in long sessions
Biphasic (bidirectional)Each pulse has a phase each way, symmetrical or asymmetricalCharge-balanced pulses leave no net charge: far less skin irritation

By pulse shape

ShapeWhat matters about it
RectangularSteepest rise: fires healthy nerve most efficiently; the standard for TENS and NMES
TriangularSlow rise and fall: healthy nerve accommodates, so it is used for partly denervated muscle
TrapezoidalA ramped rectangle: gentler rise
SinusoidalA smooth wave each way
Semi-sinusoidalHalf of a sine wave, one direction only: the shape of diadynamic currents
ExponentialA rise that starts slowly and steepens: for denervated muscle

No shape is better in itself; its rise time and the charge it carries decide what it fires.

By modulation

Nerves stop responding to an unchanging stimulus: they adapt. Modulation keeps changing the stimulus to delay that.

  • Unmodulated: steady pulses at one frequency. The book calls 80 to 150 Hz inhibitory and 1 to 50 Hz irritative.
  • Burst (group) modulation: pulses come in trains with pauses, usually for muscle stimulation.
  • Amplitude modulation: pulse strength rises and falls, usually for analgesia.
  • Frequency modulation: pulse rate changes, usually for analgesia.

By clinical use (the book's list)

CurrentMain use
Pulsed (impulse) galvanic currentAnalgesia, trophic and galvanic-like effects
Tetanic (faradic) current, 50 HzStimulation of innervated muscle
Exponential current, 0.2 to 30 HzStimulation of denervated muscle
Electroanalgesic currents: TENS, high-voltage pulsed current, microcurrent, diadynamic currentsPain relief

Parameters

the low-frequency band
up to 1,000Hz
the traditional analgesic band (100 Hz the classic)
80–150Hz
a smooth tetanic contraction
50Hz
ParameterTypical rangeWhat it changes
Pulse duration50 µs to 1 ms for nerve; 10 to 500 ms for denervated muscleShort pulses fire large fibres selectively and comfortably; long pulses are needed for muscle without its nerve
Frequency1 to 150 Hz in practiceThe pattern: twitches, summation, tetanus, a buzz
AmplitudeA few mA up to about 100 mA for short pulses; hundreds of volts for HVPCWhich fibres fire: sensory, then motor, then pain
Interpulse intervalMilliseconds to secondsTime for the membrane to recover
On and off times, rampSecondsComfort, and rest for the muscle

Physiological and therapeutic effects

EffectHow it comes aboutUsed for
AnalgesiaLarge touch fibres (Aβ) fired at 80 to 150 Hz close the spinal gate on pain fibres; low-frequency strong stimulation releases endorphins and enkephalins and engages descending inhibition from the brainstem; possibly less inflammatory mediators locallyAcute and chronic musculoskeletal pain, neuralgias
Muscle contractionMotor axons (or, if denervated, muscle fibres) fire; twitches summate and fuseStrengthening, preventing atrophy, re-education
Circulation and oedemaRhythmic contractions pump veins and lymph; the direct-current component and axon reflexes widen vesselsSwelling, poor circulation
TrophicBetter blood flow and, with monophasic currents, polarity effectsSlow-healing wounds and trophic disorders (adjunct)

The gate control theory (Melzack and Wall, 1965) is the simplest picture of electroanalgesia. In the dorsal horn of the spinal cord, signals from thick, fast touch fibres switch on small inhibitory interneurons that turn down the message from thin pain fibres (Aδ and C). Rubbing a bumped elbow works the same way. Strong low-frequency stimulation adds a second, slower mechanism: the brain's own opioids.

The currents one by one

Pulsed galvanic current (impulse galvanization)

A monophasic rectangular current: pulses of one polarity on top of, or instead of, a steady galvanic current. The book gives 100 to 150 Hz; the lecture 1 to 100 Hz with pulses of 0.1 to 10 ms. It joins galvanic effects (polarity, hyperaemia) with pulse effects (a fine "micro-massage" of tissue, analgesia), with less skin irritation than continuous galvanization. The more galvanic component, the more inhibitory; the more pulse component, the more irritative. Intensity at sensory level, 10 to 20 minutes. Technique, indications and contraindications are those of galvanization.

Tetanic (faradic) current

Short rectangular pulses (0.1 to 1 ms) at about 50 Hz, delivered in trains of a few seconds with rests between, so the muscle gives a smooth, fused (tetanic) contraction and then relaxes. The name faradic comes from Faraday's induction coil, which first produced such a current. It is used to strengthen innervated muscle and prevent disuse atrophy, not for pain. Electrodes go along the muscle, one above the other, or one over the motor point. The modern form of this current is NMES, where the parameters are covered in detail.

Exponential current, for denervated muscle

When a peripheral nerve is cut or crushed, the muscle loses its nerve: there is nothing to fire with short pulses, and the muscle fibres themselves must be fired. They need long pulses (their chronaxie is long), and they have lost most of their accommodation. A slowly rising pulse therefore fires the denervated fibres while healthy nerves and skin receptors, which accommodate, ignore it: the contraction is selective and less painful.

The book sets each parameter from electrodiagnosis, and the logic is the same for each: the worse the denervation, the more the current must be adapted.

The worse the denervationBecause
The higher the intensityThe rheobase rises
The longer the pulseThe chronaxie lengthens
The slower the rise of the pulseAccommodation is lost, so a slow rise still fires muscle but spares nerve
The longer the pause between pulsesThe refractory period lengthens
The lower the frequency: 1 to 20 Hz in partial, 0.2 to about 0.7 Hz in complete denervationFunctional lability falls
The fewer the contractions per minute (fewer than about 20 in partial denervation)The muscle tires quickly

As the nerve regrows, the pulse front is made steeper and shorter. The lecture gives pulses of 10 to 500 ms at 1 to 10 Hz, 10 to 20 minutes, electrodes along the muscle; the two ranges go together only one way round, because the longer the pulse, the fewer fit into a second (a 500 ms pulse allows at most about one a second).

Electroanalgesic currents

  • TENS, the most used of all: biphasic pulses of 50 to 400 µs at 1 to 150 Hz, worn for long periods at home. It has its own topic.
  • High-voltage pulsed current (HVPC). Very short twin-peaked monophasic pulses (about 50 to 200 µs) at 100 to 500 V (the book gives 150 V) and 1 to 120 Hz (the book gives 80 to 200 Hz for analgesia). The voltage is high, so it passes the skin easily, but each pulse is so short that the average current is tiny, so it hardly heats and hardly polarises. Used for pain, muscle re-education, oedema and, with some evidence, wound healing.
  • Microcurrent. Current in microamperes (10 to 1,000 µA, below sensation), monophasic or biphasic, at low frequencies, applied over painful areas, trigger points or acupuncture points (micro-electroanalgesia in the book). It aims at tissue repair and pain without any sensation or contraction. Evidence is weak.

Diadynamic (Bernard) currents

Pierre Bernard, a French dentist, made these in the 1950s from mains current: 50 Hz alternating current, reduced in voltage and rectified so that it flows one way only. Keeping only one half of each wave gives 50 half-waves a second (MF); keeping both halves, flipped the same way, gives 100 (DF). By alternating and modulating them he made five classic forms.

FormWhat it isEffect
DF, diphasé fixeFull-wave rectified, 100 Hz, continuousStrongest but shortest analgesia; the opening current that prepares the skin
MF, monophasé fixeHalf-wave rectified, 50 Hz, continuousCoarser, more vibrating; combined stimulating, trophic and analgesic effect
CP, courtes périodes (short periods)1 s of MF alternating with 1 s of DFMore prolonged analgesia than DF; stimulating, resorbs oedema
LP, longues périodes (long periods)MF and DF alternating over several seconds, with gradual transitionsLongest-lasting but gentlest analgesia
RS, rythme syncopéMF for 1 s, then 1 s of restThe most stimulating: for innervated muscle, not for pain
The diadynamic currents
Six pale lanes stacked down the page, each with a waveform drawn as an ink line over soft lavender humps. The top lane is a sine wave swinging above and below its dashed middle: the mains. The second keeps only the humps above the middle, with flat gaps where the lower halves were: MF. The third has a hump in every half cycle, all standing up and touching: DF. Fine dotted lines line up the halves of the three. Under them a short bracket marks one mains cycle. The lower three lanes run over seconds: CP alternates stretches of spaced humps and of packed humps; LP has stretches of spaced humps, and between them extra humps grow in the gaps and shrink away again; RS has stretches of spaced humps separated by flat stretches with none.

Point at a name or a part to light it.

Bernard made these currents from the mains. Rectifying the 50 Hz alternating current either keeps one half of each cycle (MF, 50 Hz) or turns both halves the same way (DF, 100 Hz). Alternating and interrupting the two gives CP, LP and RS. Every form flows one way only, so each carries a direct-current component, needs wet pads and can burn like galvanization. An analgesic session runs DF, then CP, then LP.

Simplified: the lower three lanes squeeze seconds into the width of the upper ones, so far fewer half-waves are drawn than flow (50 or 100 a second); every half-wave is drawn as an exact half of a sine wave at one height; how strong it is is set by the operator; sources give LP's stretches different lengths, so its lane is drawn without a time; the steady galvanic base current many devices can add under the half-waves is not drawn; with it or without it the current flows one way.

2 sources
  1. Aleksiev A. Physical Medicine and Rehabilitation for English-speaking MD students. Medical University of Sofia, 2002. Dia-dynamic (DD) current: definition, derivation and types.
  2. Aleksiev AR. Physical Medicine and Rehabilitation, lecture 1 handnotes. Medical University of Sofia, 2026. Diadynamic currents (Bernard currents).
  1. 1

    DF, about 1 minute

    Strong, quick relief and lower skin resistance.
  2. 2

    CP, a few minutes

    Longer relief; the change of rhythm delays adaptation.
  3. 3

    LP, the longest part

    The longest-lasting relief.

The lecture's analgesic sequence is about 1 minute of DF, about 5 of CP and about 10 of LP. Many devices also let a small steady galvanic "base" current run under the pulses, which adds galvanic effects and the same burn risk.

Diadynamic currents are monophasic, so they have a direct-current component: wet pads, clean skin, polarity chosen as in galvanization (the painful point usually under the cathode in Bernard's original method, the electrodes often small and round over the painful point and along the nerve). They are quickly adapted to, so each session is short and a course is usually short too, about 6 to 10 daily sessions; the forms are changed to keep them working.

Application methods

  • Electrodes: carbon rubber with wet pads for monophasic currents (they need the pad's protection), self-adhesive gel electrodes for biphasic ones.
  • Placement for pain: over or around the painful area, along the course of the nerve, over the nerve root paravertebrally (segmental), or over trigger or acupuncture points.
  • Placement for muscle: one electrode over the motor point (where the nerve enters the muscle, the point that contracts it with the least current) and one over the muscle belly or tendon, or both along the muscle; a smaller active electrode localises the contraction.
  • Placement for denervated muscle: both electrodes along the muscle belly, since there is no motor point to use.
  • Monopolar or bipolar: a small active electrode with a large one elsewhere concentrates the effect under the small one; two equal electrodes spread it between them.

Dosage

AimIntensityTimeCourse
AnalgesiaSensory level: a strong but comfortable tingle, no contraction (or a gentle twitch for low-frequency analgesia)10 to 20 minutes for clinic currents; TENS longerDaily, about 6 to 15 sessions
Innervated muscleTo a visible, palpable, tolerable contraction10 to 20 minutes, on and off times in secondsDaily or several times a week
Denervated muscleTo a visible contraction, with long pulses10 to 20 minutes, possibly several times a dayWeeks to months, reassessed by electrodiagnosis

The intensity often needs a small cut in the first minutes, as skin resistance falls under wet pads. Patients adapt within minutes to a steady analgesic current, so modulation (or changing the diadynamic form) keeps it working rather than turning it up.

Indications

Grouped by what the current does:

  • Because it relieves pain: acute and chronic musculoskeletal pain (neck, low back, osteoarthritis), sprains, bruises, myofascial pain and spasm, neuralgias and radicular pain, post-traumatic and postoperative pain.
  • Because it contracts muscle: disuse atrophy after immobilisation or surgery, weakness and inhibition, re-education; denervated muscle after peripheral nerve injury (exponential and triangular currents).
  • Because contraction and vasodilation move fluid: oedema, poor local circulation.
  • Because of polarity effects (monophasic currents): trophic problems, slow wounds (adjunct).

Contraindications

The same reasoning as for every current through the body.

ReasonContraindication
The current could reach the heart or an implantPacemaker or implanted defibrillator, other implanted stimulators; electrodes across the chest; over the carotid sinus and anterior neck; serious arrhythmias
More blood flow or contraction would spread harmMalignant tumour in the area; active bleeding; acute deep vein thrombosis (contraction could dislodge a clot); acute infection and fever; unstable fracture where contraction would move it
The skin must warn, and the patient must reportBroken or diseased skin under the electrodes; loss of sensation; a patient who cannot cooperate
Sensitive areasThe pregnant abdomen, pelvis and low back; the eyes; the head in epilepsy
Monophasic currents onlyMetal implants in the current path; the cautions of galvanization, because of the direct-current component

Check yourselfAnswer, then say how sure you were: a lucky guess still counts as a weak point.

1The same electrodes over the quadriceps deliver pulses at motor intensity, first at 2 Hz, then at 50 Hz. What does the patient's muscle do?

2A patient with an acutely painful shoulder after a strain has diadynamic therapy. Which form is used first, and why?

3Why is a slowly rising (exponential or triangular) pulse used to stimulate a denervated muscle?

4Which low-frequency current carries the greatest risk of a chemical burn under the electrode in a long session?

5A colleague says a 100 Hz current is always sensory and a 10 Hz current is always motor. What is wrong with this?

Sources

  1. Aleksiev A. Physical Medicine and Rehabilitation for English-speaking MD students. Medical University of Sofia, 2002. Treatment by low frequency impulse current (classification by polarity, shape, modulation and clinical use; impulse galvanization, tetanic current, exponential current, electroanalgesic current, high voltage); Dia-dynamic (DD) current.
  2. Aleksiev AR. Physical Medicine and Rehabilitation, lecture 1 handnotes. Medical University of Sofia, 2026. Low-frequency pulsed currents; pulsed galvanic current; tetanic current; exponential current and denervated muscle; diadynamic currents (Bernard currents); high-voltage pulsed current; microcurrent.
  3. Watson T, Nussbaum EL (eds). Electrophysical Agents, evidence-based practice, 13th edition. Elsevier, 2021.
  4. Cameron MH. Physical Agents in Rehabilitation, an evidence-based approach to practice, 5th edition. Elsevier, 2018. Electrical currents, pain control and muscle contraction.
  5. Robinson AJ, Snyder-Mackler L. Clinical Electrophysiology, electrotherapy and electrophysiologic testing, 3rd edition. Lippincott Williams and Wilkins, 2008.
  6. American Physical Therapy Association, Section on Clinical Electrophysiology. Electrotherapeutic terminology in physical therapy. APTA, 2001.
  7. Melzack R, Wall PD. Pain mechanisms, a new theory. Science, 1965.

Checked against its sources on 05 Nov 2026.